UV–Visible (UV-Vis) Spectroscopy: Principles, Instrumentation, and Applications

Ultraviolet–visible (UV-Vis) spectroscopy is one of the most widely used analytical techniques in chemistry, biology, and environmental sciences. Its popularity comes from its versatility, simplicity, and broad applicability, as it can detect nearly every molecule under the right conditions. From identifying compounds to measuring DNA, proteins, and reaction kinetics, UV-Vis spectroscopy is a foundational tool in student laboratories and research facilities alike.

This student-friendly, SEO-optimized guide explains how UV-Vis spectroscopy works, its underlying principles, instrumentation, experimental procedure, and real-world applications.


What Is UV-Vis Spectroscopy?

UV-Vis spectroscopy measures how much ultraviolet (UV) or visible light a sample absorbs. When UV-Vis light passes through a sample, some wavelengths are absorbed while others are transmitted. The amount of transmitted light is measured as transmittance (T), and absorbance (A) is calculated using the equation:

A=−log⁡(T)A = -\log(T)

By plotting absorbance against wavelength, an absorbance spectrum is obtained. Each compound produces a unique absorbance spectrum based on its chemical structure, making UV-Vis spectroscopy useful for both qualitative and quantitative analysis.


Why Is UV-Vis Spectroscopy Important?

UV-Vis spectroscopy is widely used because it can:

  • Identify compounds and functional groups

  • Measure concentration using Beer’s Law

  • Quantify DNA, RNA, and proteins

  • Monitor reaction kinetics over time

  • Serve as a detector in chromatography techniques

  • Analyze water quality and environmental samples

Although UV-Vis is not the most sensitive spectroscopic method, it remains a general-purpose analytical tool with broad applicability.


Principles of UV-Vis Spectroscopy

https://jascoinc.com/wp-content/uploads/2022/01/Figure-1-1.png
https://teaching.shu.ac.uk/hwb/chemistry/tutorials/molspec/eltrans.gif
https://upload.wikimedia.org/wikipedia/commons/7/7b/Molecule_HOMO-LUMO_diagram.svg

UV and Visible Light Regions

  • Ultraviolet (UV): 100–400 nm

  • Visible light: 400–700 nm

  • Deep UV: 100–200 nm (rarely used due to limited light sources)

Most UV-Vis spectrometers operate between 200–800 nm.


Photon Absorption and Electronic Transitions

When a photon strikes a molecule and is absorbed, an electron is promoted from a lower energy state (ground state) to a higher energy excited state. In UV-Vis spectroscopy, this involves transitions between molecular orbitals:

  • HOMO: Highest Occupied Molecular Orbital

  • LUMO: Lowest Unoccupied Molecular Orbital

The energy difference between HOMO and LUMO is called the band gap. A photon is absorbed only if its energy exactly matches this gap.


Common UV-Vis Electronic Transitions

  • π → π* (double bonds; strong absorption)

  • n → π* (non-bonding electrons)

  • σ → σ* (single bonds; deep UV, less useful)

Molecules with double bonds and conjugation absorb strongly in the UV-Vis range. Increased conjugation generally leads to greater absorbance and longer wavelengths.


Why Are UV-Vis Peaks Broad?

UV-Vis spectra often show broad peaks or shoulders due to overlapping vibrational and rotational energy levels, resulting in slightly different absorption energies.


Color and UV-Vis Absorption

A common misconception is that a compound absorbs the color it appears. In reality:

  • A compound appears colored because it transmits that color

  • It absorbs light of the complementary color

  • The wavelength of maximum absorbance (λmax) corresponds to the complementary color

For example:

  • A blue dye absorbs in the orange/red region

  • A red dye absorbs in the green region

A color wheel is often used to predict absorbance behavior.


Beer’s Law and Quantitative Analysis

Absorption follows Beer’s Law:

A=εbCA = \varepsilon b C

Where:

  • A = absorbance

  • ε = molar attenuation coefficient

  • b = path length (usually 1 cm)

  • C = concentration

The molar attenuation coefficient is a property of the molecule and depends on functional groups and conjugation. If absorbance is low, molecules can sometimes be chemically tagged to enhance detection.


UV-Vis Instrumentation

https://jascoinc.com/wp-content/uploads/2022/01/Figure-1-1.png
https://www.ssi.shimadzu.com/sites/ssi.shimadzu.com/files/products/images/spectroscopy/faq/uv-vis-faq-instrument-design-spectrophotometer-function.png
https://www.ssi.shimadzu.com/sites/ssi.shimadzu.com/files/products/images/spectroscopy/faq/uv-vis-faq-instrument-design-diode-array-spectrophotometer.png
Light Sources
  • Deuterium lamp: UV region (170–375 nm)

  • Tungsten filament lamp: Visible region (350–2,500 nm)


Wavelength Selection

  • Filters: simple wavelength selection

  • Monochromators: spatially separate wavelengths and scan spectra

  • Diode-array detectors: capture full spectra simultaneously

Diode-array instruments are faster but more complex and expensive.


UV-Vis Experimental Procedure

1. Calibrate the Spectrometer

  • Turn on the instrument and allow lamps to warm up (~20 minutes)

  • Fill a cuvette with solvent (blank)

  • Clean the cuvette exterior

  • Measure the blank to correct for solvent absorption


2. Collect an Absorbance Spectrum

  • Rinse cuvette with sample and fill ~¾ full

  • Insert cuvette correctly and block ambient light

  • Scan from 200–800 nm

  • Determine λmax

  • Repeat scans to estimate error

To build a calibration curve, measure spectra at different concentrations. Absorbance values above 1.5 should be avoided—dilute samples if necessary.


3. Reaction Kinetics Using UV-Vis

UV-Vis is ideal for studying reaction rates:

  • Record initial absorbance

  • Add reagent to initiate reaction

  • Measure absorbance at λmax over time

  • Convert absorbance to concentration using Beer’s Law

  • Determine reaction order and rate constants


Results and Interpretation

https://www.researchgate.net/publication/46262542/figure/fig4/AS%3A340546762625028%401458204083530/UV-VIS-absorption-spectrum-of-acid-blue-3-dye-in-water.png
https://www.researchgate.net/publication/333413404/figure/fig4/AS%3A767750154760192%401560057309043/a-Time-dependent-UV-Vis-spectrum-b-the-pseudo-first-order-kinetics-plot-for.png
https://www.researchgate.net/publication/331661628/figure/fig2/AS%3A11431281272474763%401724104783118/Absorbance-spectra-of-food-dyes-in-the-red-violet-color-range-A-Schematic-diagram-of-a.tif
  • Blue dyes absorb in the orange/red region

  • Red dyes absorb in the green region

  • Kinetics plots often show exponential decay, indicating first-order reactions


Applications of UV-Vis Spectroscopy

https://cellculture2.altervista.org/wp-content/uploads/2024/02/image-21.png
https://www.researchgate.net/publication/260386871/figure/fig3/AS%3A339892933546011%401458048198454/Figure-3-UV-absorption-spectra-of-plasmid-DNA-at-260-nm-in-the-presence-of-an.png
https://www.researchgate.net/publication/51873414/figure/fig7/AS%3A667216731205632%401536088273020/UV-spectra-of-cytochrome-c-typical-spectra-of-cytochrome-c-with-three-absorbance-peaks-of.ppm
Biological Applications
  • Proteins: absorb strongly at 280 nm

  • DNA/RNA: absorb at 260 nm

  • Purity assessment: A260/A280 ratio


Chemical and Analytical Applications

  • Compound identification

  • Concentration determination

  • Water and environmental analysis

  • Chromatography detection

  • Multi-component analysis with diode-array detectors


Advantages and Limitations of UV-Vis Spectroscopy

Advantages

  • Simple and fast

  • Non-destructive

  • Applicable to many compounds

  • Ideal for teaching laboratories

Limitations

  • Lower sensitivity than fluorescence

  • Limited selectivity

  • Requires UV-absorbing chromophores


Conclusion

UV-Vis spectroscopy is a fundamental analytical technique that plays a critical role in chemistry, biology, and environmental science. By measuring light absorption and applying Beer’s Law, students can identify compounds, determine concentrations, and study reaction kinetics. While not the most sensitive technique, UV-Vis remains one of the most versatile and widely used tools in the laboratory.

Understanding UV-Vis spectroscopy provides a strong foundation for more advanced analytical methods and real-world scientific applications.

Alisha G C

Alisha G C is an MBBS student at Nepalgunj Medical College, Banke, Nepal. She writes biology notes at www.thesciencenotes.com. https://www.nature.com/articles/d41586-025-00589-z

Leave a Reply

Your email address will not be published. Required fields are marked *